head_banner

Portable and Hospital use anesthesia syringe pump

Portable and Hospital use anesthesia syringe pump

Short Description:

Features

1. Work mode:

constant infusion, intermittent infusion, TCI (Target Control Infusion).

2. Multiply infusion mode:

easy mode, flow rate, time, body weight, plasma TCI, effect TCI

3. TCI calculate mode:

maximum mode, increment mode, constant mode.

4. Compatible with syringe of any standard.

5. Adjustable bolus rate 0.1-1200 ml/h in 0.01, 0.1, 1, 10 ml/h increment.

6. Adjustable KVO rate 0.1-1 ml/h in 0.01 ml/h increments.

7. Automatic anti-bolus.

8. Drug library.

9. History log of 50,000 events.

10. Stackable for multiple channels.


Product Detail

Product Tags

Portable and Hospital use anesthesia syringe pump,
Anesthesia Syringe Pump,
Specifications

Model KL-605T
Syringe Size 5, 10, 20, 30, 50/60 ml
Applicable Syringe Compatible with syringe of any standard
VTBI 1-1000 ml (in 0.1, 1, 10 ml increments)
Flow Rate Syringe 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)

Syringe 10 ml: 0.1-300 ml/h

Syringe 20 ml: 0.1-600 ml/h

Syringe 30 ml: 0.1-800 ml/h

Syringe 50/60 ml: 0.1-1200 ml/h

Bolus Rate 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)

10 ml: 0.1-300 ml/h

20 ml: 0.1-600 ml/h

30 ml: 0.1-800 ml/h

50/60 ml: 0.1-1200 ml/h

Anti-Bolus Automatic
Accuracy ±2% (mechanical accuracy≤1%)
Infusion Mode 1. Easy mode

2. Flow rate

3. Time-based

4. Body weight

5. Plasma TCI

6. Effect TCI

KVO Rate 0.1-1 ml/h (in 0.01 ml/h increments)
Alarms Occlusion, near empty, end program, low battery, end battery,

AC power off, motor malfunction, system malfunction, standby,

pressure sensor error, syringe installation error, syringe drop off

Additional Features Real-time infused volume, automatic power switching,

automatic syringe identification, mute key, purge, bolus, anti-bolus,

system memory, history log

Drug Library Available
Occlusion Sensitivity High, medium, low
History Log 50000 events
Power Supply, AC 110-230 V, 50/60 Hz, 20 VA
Battery 14.8 V, rechargeable
Battery Life 8 hours at 5 ml/h
Working Temperature 5-40℃
Relative Humidity 20-90%
Atmospheric Pressure 700-1060 hpa
Size 245*120*115 mm
Weight 2.5 kg
Safety Classification Class Ⅱ, type BF

Model
KL-605T
Syringe Size : 5, 10, 20, 30, 50/60 ml
Applicable Syringe : Compatible with syringe of any standard
VTBI :1-1000 ml (in 0.1, 1, 10 ml increments)
Flow Rate
Syringe 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)
Syringe 10 ml: 0.1-300 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-800 ml/h
Syringe 50/60 ml: 0.1-1200 ml/h
Bolus Rate
5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)
10 ml: 0.1-300 ml/h
20 ml: 0.1-600 ml/h
30 ml: 0.1-800 ml/h
50/60 ml: 0.1-1200 ml/h
Anti-Bolus :Automatic
Accuracy :±2% (mechanical accuracy≤1%)
Infusion Mode
1. Easy mode
2. Flow rate
3. Time-based
4. Body weight
5. Plasma TCI
6. Effect TCI
KVO Rate : 0.1-1 ml/h (in 0.01 ml/h increments)
Alarms:
Occlusion, near empty, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarm,
pressure sensor error, syringe installation error, syringe drop off
Additional Features :
Real-time infused volume, automatic power switching,
automatic syringe identification, mute key, purge, bolus, anti-bolus,
system memory, history log
Drug Library :Available
Occlusion Sensitivity: High, medium, low
Power Supply, AC :110/230 V, 50/60 Hz, 20 VA
Battery: 14.8 V, rechargeable
Battery Life: 8 hours at 5 ml/h
Working Temperature : 5-40℃
Relative Humidity :20-90%
Atmospheric Pressure :700-1060 hpa
Size :245*120*115 mm
Weight :2.5 kg
Safety Classification :Class Ⅱ, type BF