Q: Is the pump open system?
A: Yes, Universal IV set can be used with our Infusion Pump after calibration.
Q: Does the pump compatible with Micro IV Set (1 ml=60 drops)?
A: Yes, all of our pumps are compatible with IV Set of 15/20/60 dorps.
Q: Is it a peristaltic pumping mechanism?
A: Yes, curvilinear peristaltic.
Q: What is the difference between PURGE and BOLUS function?
A: Purge used to remove air in line before infusion. Bolus can be administered for infusion therapy during infusion. Both purge and bolus rate are programmable.
Specifications
Model | ZNB-XAII |
Pumping Mechanism | Curvilinear peristaltic |
IV Set | Compatible with IV sets of any standard |
Flow Rate | 1-1500 ml/h (in 0.1 ml/h increments) |
Purge, Bolus | 100-1500 ml/h (in 0.1 ml/h increments)
Purge when pump stops, bolus when pump starts |
Accuracy | ±3% |
*Inbuilt Thermostat | 30-45℃, adjustable |
VTBI | 1-20000 ml (in 0.1 ml increments) |
Infusion Mode | ml/h, drop/min, time-based, body weight, nutrition |
KVO Rate | 0.1-5 ml/h (in 0.1 ml/h increments) |
Alarms | Occlusion, air-in-line, door open, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, standby |
Additional Features | Real-time infused volume, automatic power switching, mute key,
purge, bolus, system memory, history log, key locker, drug library, rotary knob, change flow rate without stopping the pump |
Drug Library | Available |
Occlusion Sensitivity | High, medium, low |
History Log | 50000 events |
Air-in-line Detection | Ultrasonic detector |
Wireless management | Optional |
Drop Sensor | Optional |
Vehicle Power (Ambulance) | 12±1.2 V |
Power Supply, AC | 110/230 V (optional), 50-60 Hz, 20 VA |
Battery | 9.6±1.6 V, rechargeable |
Battery Life | 5 hours at 25 ml/h |
Working Temperature | 10-30℃ |
Relative Humidity | 30-75% |
Atmospheric Pressure | 860-1060 hpa |
Size | 130*145*228 mm |
Weight | 2.5 kg |
Safety Classification | Class Ⅰ, type CF |
Product detail pictures:

Related Product Guide:
That has a positive and progressive attitude to customer's interest, our organization consistently improves our products quality to satisfy the demands of shoppers and further focuses on safety, reliability, environmental specifications, and innovation of OEM/ODM China Elastomeric Infusion Pump - ZNB-XAII Infusion Pump – KellyMed , The product will supply to all over the world, such as: Comoros, Venezuela, Seattle, We are sincerely looking forward to cooperate with customers all over the world. We believe we can satisfy you with our high-quality products and perfect service . We also warmly welcome customers to visit our company and purchase our products.