Specification for Rotary Dual Channel Enteral Feeding Pump with Automatic Flush Function
Model | KL-5051N |
Pumping Mechanism | Rotary with automatic flush function |
Enteral Feeding Set | Compatible with T-shaped enteral feeding set, double channel |
Flow Rate | 1-2000 ml/h (in 0.1 ml/h increments) |
Suck/Flush Rate | 100~2000ml/h(in 1 ml/h increments) |
Purge/Bolus Volume | 1-100 ml (in 1 ml increments) |
Suck/Flush Rate | 100-2000 ml/h (in 1 ml/h increments) |
Suck/Flush Volume | 1-1000 ml (in 1 ml increments) |
Accuracy | ±5% |
VTBI | 1-20000 ml (in 0.1 ml increments) |
Feeding Mode | Continuous, Intermittent, Pulse, Time, Scientific. Flush |
KTO | 1-10 ml/h (in 0.1 ml/h increments) |
Alarms | occlusion, air-in-line, low battery, end battery, AC power off, tube error, rate error, motor error, hardware error, over temperature, standby, sleeping. |
Additional Features | Real-time infused volume, automatic power switching, mute key, purge, bolus, system memory, history log, key locker, suck, cleaning |
*Fluid Warmer | Optional (30-37℃, over temperature alarm) |
Occlusion Sensitivity | 3 Levels: High、middle、low |
Air-in-line Detection | Ultrasonic detector |
History Log | 30 days |
Wireless management | Optional |
Power Supply, AC | 110-240V, 50/60HZ, ≤100VA |
Vehicle Power (Ambulance) | 24V |
Battery | 12.6 V, rechargeable, Lithium |
Battery Life | 5 hours at 125ml/h |
Working Temperature | 5-40℃ |
Relative Humidity | 10-80% |
Atmospheric Pressure | 860-1060 hpa |
Size | 126(L)*174(W)*100(H) mm |
Weight | 1.6 kg |
Safety Classification | Class Ⅱ, type BF |
Fluid Ingress Protection | IP23 |
Product detail pictures:
Related Product Guide:
Our business aims to operating faithfully, serving to all of our clients , and working in new technology and new machine continuously for Good quality Kangaroo Feeding Pump Bags - Dual Feeding Pump with Automatic Flush Function Enteral Nutrition Pump use in ICU KL-5051N – KellyMed , The product will supply to all over the world, such as: UAE, Czech republic, Juventus, As the world economic integration bringing challenges and opportunities to the xxx industry, our company , by carrying on our teamwork, quality first, innovation and mutual benefit, are confident enough to provide our clients sincerely with qualified products, competitive price and great service, and to build a brighter future under the spirit of higher, faster, stronger with our friends together by carrying on our discipline.
By ron gravatt from Jersey - 2017.09.09 10:18
The company can keep up with the changes in this industry market, product updates fast and the price is cheap, this is our second cooperation, it's good.
By Dawn from Vancouver - 2017.02.18 15:54
Write your message here and send it to us